Common questions may occur when using ultrasonic cement analyzer
1. The pressure does not rise
(1) Check whether the electromagnetic valve at the front of the booster pump is working properly; (2) whether the air source pressure is normal; (3) whether there is any leakage in the air path; (4) whether the seal of the kettle caps is normal.
2. The acoustic time is undetected
(1) Check whether the software connection is normal; (2) whether the BNC signal transmission line is normal; (3) whether the probe is normal; (4) whether the coupling agent is normal; (5) If there is no problem with the above, please contact us.
3. Abnormal acoustic time intensity data
Use the standard calibration block to recalibrate and compare the experimental data.
4. Replacement of the wedge seal O-rings of the slurry cup's caps
The rubber O-rings need to be replaced when tarnish and deteriorate after high-temperature testing. The wedge-shaped sealing O-rings need to be replaced when its inner diameter becomes distinctly larger and the sealing effect cannot be achieved.
Technical Parameters
Working Temperature |
204℃ |
Working Pressure |
138MPa |
Input Voltage |
AC220V 50Hz/110V /60Hz |
Heater Power |
2000W |
Input Power |
2100W |
Air Supply Pressure |
0.4-0.7MPa |
Dimensions (W*D*H) |
50*70*100cm |
Weight |
150kg |
Plant Strength